Instrumentation and Measurement for the Junior Engineer Papers: Errors, Indicating Instruments, Bridges, Transducers and Process Control
2. Indicating instruments, shunts, multipliers and the energy meter
| Type | Works on | Scale and use |
|---|---|---|
| PMMC (moving coil) | DC only | Uniform scale, high accuracy, low power use; with a rectifier it reads AC |
| Moving iron | AC and DC | Non-uniform, cramped scale; rugged and cheap; reads RMS |
| Dynamometer | AC and DC | Used as the wattmeter, which measures real power |
| Induction | AC only | The domestic energy meter, which counts kilowatt-hours |
A moving-coil galvanometer of full-scale current 1 mA and resistance 50 Ω becomes a 1 A ammeter with a low shunt in parallel, R_sh = I_g R_g ÷ (I − I_g) = 0.001 × 50 ÷ 0.999 ≈ 0.05 Ω. It becomes a 100 V voltmeter with a high multiplier resistor in series, R_s = V ÷ I_g − R_g = 100 ÷ 0.001 − 50 = 99,950 Ω. An ammeter has very low resistance and goes in series; a voltmeter has very high resistance and goes in parallel. An energy meter marked 600 revolutions per kWh turns 60 times when a 1 kW load runs for 6 minutes (0.1 kWh).
A moving-coil galvanometer of full-scale current 1 mA and resistance 50 Ω becomes a 1 A ammeter with a low shunt in parallel, R_sh = I_g R_g ÷ (I − I_g) = 0.001 × 50 ÷ 0.999 ≈ 0.05 Ω. It…
3. Bridges and transducers
A Wheatstone bridge is balanced when no current flows in the detector, which happens when the ratios of the arms are equal: if R₁ = 50 Ω and R₂ = 100 Ω form one side and R₃ = 200 Ω is the known arm, the unknown arm is R_x = R₂R₃ ÷ R₁ = 400 Ω. Related bridges are the Kelvin (low resistance), Maxwell (inductance) and Schering (capacitance and dielectric loss). A strain gauge changes resistance with strain, with gauge factor GF = (ΔR ÷ R) ÷ strain: for GF = 2 and a strain of 500 microstrain the change is 0.1%, so a 120 Ω gauge changes by 0.12 Ω, which a bridge converts to a voltage.
| Transducer | Measures | Principle and note |
|---|---|---|
| Thermocouple | Temperature | Seebeck effect: a voltage where two metals join; type K is chromel-alumel, type J iron-constantan |
| RTD, Pt100 | Temperature | Resistance of platinum rises with temperature: 100 Ω at 0 °C and about 138.5 Ω at 100 °C |
| Thermistor | Temperature | Semiconductor; an NTC thermistor's resistance falls as temperature rises |
| LVDT | Displacement | Differential transformer: output is zero at the centre and its phase shows the direction |
| Piezoelectric | Dynamic force, pressure, vibration | Charge on quartz when stressed; cannot read a steady (static) value |
A Wheatstone bridge is balanced when no current flows in the detector, which happens when the ratios of the arms are equal: if R₁ = 50 Ω and R₂ = 100 Ω form one side and R₃ = 200 Ω is the…
4. Pressure, flow, level and the control loop
Pressure is measured by the manometer, the Bourdon tube (the curved tube that straightens under pressure, the usual gauge), the diaphragm and bellows, and electrically by strain-gauge and piezoelectric sensors. Flow is measured by differential-pressure devices (the orifice plate and venturi) with Q ∝ √ΔP, so a fourfold rise in the differential pressure doubles the flow; by the rotameter (a float in a tapered tube, variable area); and by electromagnetic and ultrasonic meters, which have no moving parts. Level uses floats, differential-pressure cells, capacitance probes and ultrasonics.
In process control the standard transmitter signal is 4 to 20 mA, whose 'live zero' of 4 mA distinguishes a zero reading from a broken wire. A 0 to 200 °C transmitter gives 12 mA at (12 − 4) ÷ 16 × 200 = 100 °C and 8 mA at 50 °C. Controller modes: on-off (simple, oscillates), proportional (leaves a steady offset), PI (the integral action removes the offset), and PID (the derivative action anticipates the change and improves stability).
Pressure is measured by the manometer, the Bourdon tube (the curved tube that straightens under pressure, the usual gauge), the diaphragm and bellows, and electrically by strain-gauge and…
Worked Numericals: Shunt, Multiplier, Bridge and Gauge Factor
| Problem | Working | Result |
|---|---|---|
| Extend a 1 mA, 100 Ω meter to read 100 mA | R_sh = R_m I_m/(I − I_m) = 100 × 1/99 | ≈ 1.01 Ω in parallel |
| Make the same meter read 10 V | R_se = V/I_m − R_m = 10/0.001 − 100 | 9900 Ω in series |
| Strain gauge with ΔR/R = 0.002 at strain 0.001 | gauge factor = (ΔR/R)/strain | 2 |
| Wheatstone bridge balanced with P = 100 Ω, Q = 1000 Ω, R = 250 Ω | S = QR/P | 2500 Ω |
Measurement numericals
Transducers: What Each One Senses
| Transducer | Measures | Principle |
|---|---|---|
| Strain gauge | strain, force or pressure | resistance changes with length and cross-section |
| LVDT | displacement | a movable core changes the mutual inductance between windings |
| Thermocouple | temperature | Seebeck effect: a junction of two metals gives an emf |
| Piezoelectric crystal | force, vibration | a crystal under stress produces a charge |
| Thermistor | temperature | resistance changes strongly with temperature |
Transducers and their principle
Second Pass: Shunts, Limiting Error, Wattmeter Reading and Transmitter Scaling
A shunt in parallel with a galvanometer takes the extra current: Rsh = Ig Rg/(I − Ig). A series multiplier gives Rs = V/Ig − Rg. A limit of error quoted on full scale becomes a larger percentage at a low reading. A wattmeter reads V I cos φ, and a 4 to 20 mA signal spans its range linearly with 4 mA as the zero.
| Item | Working | Answer |
|---|---|---|
| 1 mA, 50 Ω galvanometer to 1 A | 0.001 × 50/0.999 | Rsh ≈ 0.05 Ω |
| Same meter to 50 V | 50/0.001 − 50 | Rs = 49,950 Ω |
| ±1 V on 100 V scale, reading 80 V | 1/80 | 1.25% of reading |
| 230 V, 10 A, pf 0.5 | 230 × 10 × 0.5 | 1,150 W |
A shunt in parallel with a galvanometer takes the extra current: Rsh = Ig Rg/(I − Ig). A series multiplier gives Rs = V/Ig − Rg. A limit of error quoted on full scale becomes a larger…
Key takeaways
- Accuracy is closeness to truth, precision is repeatability; a full-scale percentage error is fixed in volts, so read in the upper part of the scale.
- PMMC reads DC only on a uniform scale, moving iron reads AC and DC on a cramped scale, the dynamometer is the wattmeter and the induction type is the energy meter.
- An ammeter needs a low-resistance shunt in parallel; a voltmeter needs a high-resistance multiplier in series, R_s = V ÷ I_g − R_g.
- A balanced Wheatstone bridge gives R_x = R₂R₃ ÷ R₁; gauge factor is (ΔR ÷ R) ÷ strain; Pt100 is 100 Ω at 0 °C.
- Differential-pressure flow varies as √ΔP; the 4 to 20 mA signal has a live zero; PI removes offset and PID adds anticipation.
- A shunt (R_m I_m/(I − I_m)) is in parallel and extends the current range; a multiplier (V/I_m − R_m) is in series and extends the voltage range.
- An LVDT senses displacement, a thermocouple temperature through the Seebeck effect, and a piezoelectric crystal force.
Practice questions (25)
Attempt each one before opening the answer. Every explanation names the tempting wrong option as well as the right one, because that is where marks are lost.
A voltmeter of range 0 to 100 V is accurate to 1% of full scale. For a reading of 40 V the possible error, as a percentage of the reading, is:
Show answer
Answer: A — 2.5%
The error is 1% of 100 V = ±1 V at every point, and 1 V ÷ 40 V = 2.5% of the reading. The error is not 1% of the reading, which is why a reading at the top of the scale is more reliable than one near the bottom.Which instrument can measure only direct current, has a uniform scale and a permanent magnet?
Show answer
Answer: A — PMMC (moving coil)
The permanent-magnet moving-coil instrument deflects in proportion to the current, giving a uniform scale, and on AC the pointer can only vibrate about zero, so it reads DC only unless a rectifier is added. Moving-iron and dynamometer types work on both AC and DC and have non-uniform scales.A galvanometer with full-scale current 1 mA and resistance 50 Ω is to be used as a voltmeter of range 0 to 100 V. The series resistance required is:
Show answer
Answer: A — 99,950 Ω
The total resistance must be V ÷ I_g = 100 ÷ 0.001 = 100,000 Ω, and the meter already supplies 50 Ω, so the series multiplier is 100,000 − 50 = 99,950 Ω. The figure 100,000 Ω forgets to subtract the coil resistance.
Show the other 22 questions
An energy meter is marked 600 revolutions per kWh. A load of 1 kW runs for 6 minutes. The number of revolutions made by the disc is:
Show answer
Answer: A — 60
Six minutes is 0.1 hour, so the energy is 1 kW × 0.1 h = 0.1 kWh, and the disc turns 0.1 × 600 = 60 times. The answer 360 treats the six minutes as 6 × 60 seconds without converting to hours.A Wheatstone bridge has R₁ = 50 Ω and R₂ = 100 Ω in the ratio arms, and a known arm R₃ = 200 Ω. At balance the unknown arm R_x, with R_x ÷ R₃ = R₂ ÷ R₁, is:
Show answer
Answer: A — 400 Ω
R_x = R₃ × R₂ ÷ R₁ = 200 × 100 ÷ 50 = 400 Ω. Inverting the ratio gives 100 Ω, which is the usual slip, so check which arm sits opposite which before substituting.A strain gauge of resistance 120 Ω and gauge factor 2 is subjected to a strain of 500 microstrain (500 × 10⁻⁶). The change in its resistance is:
Show answer
Answer: A — 0.12 Ω
ΔR ÷ R = GF × strain = 2 × 500 × 10⁻⁶ = 0.001, so ΔR = 120 × 0.001 = 0.12 Ω. The tiny change is why a bridge circuit and an amplifier are needed to read a strain gauge.In a differential-pressure flow meter (an orifice plate), the differential pressure rises to four times its original value. The flow rate becomes:
Show answer
Answer: A — twice
Flow is proportional to the square root of the differential pressure, and √4 = 2, so the flow doubles. Saying four times treats the flow as proportional to the pressure, which would be true only for a linear device.A 4 to 20 mA transmitter is calibrated for 0 to 200 °C. A reading of 8 mA corresponds to a temperature of:
Show answer
Answer: A — 50 °C
The span of the current is 16 mA, so (8 − 4) ÷ 16 = 25% of the range, which is 25% of 200 = 50 °C. Treating 4 mA as zero current gives 8 ÷ 20 × 200 = 80 °C, ignoring the live zero, and 12 mA is the midpoint, 100 °C.A 1 mA, 100 Ω galvanometer is to be converted into a voltmeter of range 10 V. The series resistance needed is:
Show answer
Answer: A — 9900 Ω
The total resistance must be V/I = 10 000 Ω, of which the meter supplies 100 Ω, so 9900 Ω is added in series. The value 10 000 Ω ignores the meter's own resistance.A thermocouple measures temperature using the:
Show answer
Answer: A — Seebeck effect
A junction of two dissimilar metals produces an emf that depends on temperature, the Seebeck effect. The Hall effect relates to magnetic fields and the piezoelectric effect to crystals under stress.A galvanometer of full-scale current 1 mA and resistance 50 Ω is to be used as an ammeter of range 0 to 1 A. The shunt resistance required is nearest to:
Show answer
Answer: A — 0.05 Ω
The shunt carries 0.999 A at 50 mV across the meter, so Rsh = 0.05/0.999 ≈ 0.05 Ω. A series resistor, not a shunt, would be 49,950 Ω for a 50 V voltmeter.The same 1 mA, 50 Ω galvanometer is to be converted into a voltmeter of range 0 to 50 V. The series multiplier is:
Show answer
Answer: A — 49,950 Ω
Total resistance = 50 V/1 mA = 50,000 Ω, and subtracting the meter's own 50 Ω gives 49,950 Ω.A voltmeter of range 0 to 100 V is guaranteed accurate to ±1 V. For a reading of 80 V, the limiting error as a percentage of the reading is:
Show answer
Answer: A — 1.25%
Error as a percentage of reading = 1/80 × 100 = 1.25%. The 1% is the percentage of full scale.A wattmeter is connected to a single-phase load drawing 10 A from a 230 V supply at a power factor of 0.5. The wattmeter reads:
Show answer
Answer: A — 1,150 W
A wattmeter reads real power V I cos φ = 230 × 10 × 0.5 = 1,150 W. The 2,300 is the apparent power in VA.Consider the following statements:
1. The resistance of an NTC thermistor falls as temperature rises.
2. The resistance of a platinum RTD rises as temperature rises.
Which is/are correct?Show answer
Answer: A — Both 1 and 2
A semiconductor thermistor has a negative temperature coefficient, while a metal such as platinum has a positive coefficient. Both statements are correct.A strain gauge of 120 Ω changes resistance by 0.24 Ω under a strain of 1,000 microstrain. Its gauge factor is:
Show answer
Answer: A — 2
GF = (ΔR/R)/strain = (0.24/120)/0.001 = 0.002/0.001 = 2, a typical value for metallic foil gauges.A 4 to 20 mA transmitter output reads 12 mA. The measured variable is at what fraction of its span?
Show answer
Answer: A — 50%
Fraction = (12 − 4)/(20 − 4) = 8/16 = 50%. The 60% divides 12 by 20, forgetting the live zero at 4 mA.A moving-coil meter has a full-scale current of 50 µA. Its sensitivity as a voltmeter is:
Show answer
Answer: A — 20 kΩ per volt
Sensitivity = 1 ÷ full-scale current = 1 ÷ 50 µA = 20,000 Ω/V. A 10 V range then has an input resistance of 200 kΩ. 50 kΩ/V is the tempting misreading of the 50.A linear variable differential transformer (LVDT) is used to measure:
Show answer
Answer: D — Linear displacement
An LVDT's output voltage is proportional to the displacement of its core, with a null at the centre and no sliding contact. A thermocouple is for temperature, the tempting wrong option.Which of the following is the correct match of an AC bridge with its use?
Show answer
Answer: A — Maxwell bridge — inductance
The Maxwell bridge measures inductance, the Schering bridge measures capacitance and dielectric loss, and the Wien bridge measures frequency or serves in an oscillator. Schering for inductance is the tempting mix-up.A Type K thermocouple is made of which pair of metals?
Show answer
Answer: A — Chromel and alumel
Type K is chromel (nickel-chromium) against alumel (nickel-aluminium). Iron-constantan is type J and copper-constantan is type T, the tempting neighbours.A Pt100 resistance temperature detector has a resistance of:
Show answer
Answer: A — 100 Ω at 0 °C
Pt100 means platinum with 100 Ω at 0 °C, rising by about 0.385 Ω per °C. The 100 °C reading would be about 138.5 Ω.Consider the following statements about a PID controller:
1. Integral action removes the steady-state offset.
2. Derivative action responds to the rate of change of the error and anticipates it.
Which of the statements given above is/are correct?Show answer
Answer: A — Both 1 and 2
Integral action keeps acting until the error is zero, and derivative action anticipates a changing error. Saying proportional action alone removes offset is the tempting error.A multimeter with a sensitivity of 20 kΩ per volt is set on its 10 V range. Its input resistance is:
Show answer
Answer: D — 200 kΩ
Input resistance = sensitivity × range = 20 kΩ/V × 10 V = 200 kΩ. Quoting 20 kΩ forgets to multiply by the range.Assertion (A): A 4–20 mA signal is preferred to a 0–20 mA signal in process instrumentation.
Reason (R): The live zero at 4 mA lets a broken wire (0 mA) be distinguished from a true minimum reading.
Which one of the following is correct?Show answer
Answer: A — Both A and R are true, and R is the correct explanation of A
With a live zero, 0 mA signals a fault rather than a reading, so R explains A. The ability to power the transmitter from the loop is a further benefit but not the reason given.