Measurements at Working Depth: Uncertainty and Propagation, Bridge Balance Equations, Q-Meter, Power and Instrument Transformers, True RMS, Counters, Oscilloscope and Energy Meters
Bridges: Balance Equations for Wheatstone, Kelvin, Maxwell, Anderson, Schering and Wien, and the Q-Meter
A bridge is balanced when the detector reads zero, i.e. when the products of opposite-arm impedances are equal: Z₁Z₄ = Z₂Z₃ (as complex numbers, so both magnitude and phase conditions apply — two independent adjustments are needed for an AC bridge). Wheatstone: R_x = R₂R₃/R₁ — with R₁ = 100 Ω, R₂ = 1 kΩ and R₃ = 235.0 Ω, R_x = 2350 Ω. Off balance, a small change ΔR in one equal arm R gives V₀ = E ΔR/(4R): E = 5 V, R = 120 Ω, ΔR = 0.24 Ω gives 2.5 mV (the strain-gauge bridge). Kelvin double bridge measures low resistance: a second pair of ratio arms (p, q) is connected through the link of resistance r between R_x and the standard S; with p/q = P/Q the lead and contact resistance cancel and R_x = S·P/Q — for S = 0.050 Ω and P/Q = 100/1000, R_x = 5.0 mΩ. Maxwell (inductance in terms of a capacitor in parallel with R₄): L_x = R₂R₃C₄, R_x = R₂R₃/R₄; with R₂ = 1 kΩ, R₃ = 500 Ω, R₄ = 2 kΩ, C₄ = 0.1 µF: L_x = 50 mH, R_x = 250 Ω, and Q = ωL_x/R_x = ωC₄R₄ = 1.257 at 1 kHz — frequency-dependent only through Q, so Maxwell's bridge suits low-Q coils (Q < 10) since high Q would need an impractically large R₄.
Anderson bridge handles a wide Q range: the inductor is balanced against a standard capacitor through an extra resistor branch; its balance equations are independent of frequency, at the cost of five adjustable elements and a more tedious balance. Schering bridge measures capacitance and dielectric loss: C_x = C₂R₄/R₃, R_x = R₃C₄/C₂, and tan δ = ωC_xR_x = ωC₄R₄; with C₂ = 100 pF (standard), R₃ = 500 Ω, R₄ = 1 kΩ, C₄ = 1 nF: C_x = 200 pF, R_x = 5000 Ω, tan δ at 50 Hz = 3.14 × 10⁻⁴, the standard high-voltage insulation test. Wien bridge for frequency: with equal R and C in the series and parallel arms, balance needs R₃/R₄ = 2 and f = 1/(2πRC) — 1001 Hz for 10 kΩ and 15.9 nF (it is also the frequency-selective network of the Wien oscillator). The Q-meter resonates a coil with a calibrated variable capacitor: at resonance Q = V_C/V_in (the voltage magnification), so V_in = 10 mV and V_C = 1.2 V give Q = 120, and the coil inductance follows from L = 1/(ω²C): at 1 MHz with C = 253.3 pF, L = 100 µH. Megohm (insulation) measurement uses a high-voltage source, a guard terminal that diverts surface leakage around the meter, and a long charging time.
| Bridge | Measures | Balance equations |
|---|---|---|
| Wheatstone | Medium R | R_x = R₂R₃/R₁ |
| Kelvin | Low R (milliohms) | R_x = S·P/Q |
| Maxwell | L, low Q | L_x = R₂R₃C₄, R_x = R₂R₃/R₄ |
| Schering | C and tan δ | C_x = C₂R₄/R₃, tan δ = ωC₄R₄ |
| Wien | f (or C) | f = 1/(2πRC), R₃/R₄ = 2 |
A bridge is balanced when the detector reads zero, i.e. when the products of opposite-arm impedances are equal: Z₁Z₄ = Z₂Z₃ (as complex numbers, so both magnitude and phase conditions apply…
Measurement of Voltage, Current and Power; Instrument Transformers
DC and single-phase. A moving-coil meter reads DC (and the average of rectified AC); extending range uses a series multiplier R_s = (V/I_m) − R_m for a voltmeter and a parallel shunt R_sh = I_mR_m/(I − I_m) for an ammeter; a meter loads the circuit (a voltmeter of 1 kΩ/V on a 10 V range is 10 kΩ and reads a 1 kΩ–1 kΩ divider low). A wattmeter (electrodynamic) reads VI cos φ — its current coil in series, voltage coil in parallel — and for three-phase three-wire loads the two-wattmeter method gives total power P = W₁ + W₂ regardless of balance, and for a balanced load tan φ = √3(W₁ − W₂)/(W₁ + W₂), reactive power Q = √3(W₁ − W₂). With W₁ = 2000 W and W₂ = 1000 W: P = 3000 W, tan φ = 0.577, φ = 30°, power factor 0.866 lagging, Q = 1732 var. At unity pf the two readings are equal; at pf = 0.5 one reads zero; below 0.5 it reads negative (the pressure-coil connection must be reversed).
Instrument transformers extend AC ranges and isolate the meter. A current transformer (CT) steps current down in the ratio of turns (100/5 A has nominal ratio K_n = 20); it works with a near-short-circuit secondary, so its secondary must never be opened while the primary carries current (the unopposed primary ampere-turns saturate the core and induce a dangerous voltage). A potential transformer (PT/VT) steps voltage down (11 kV/110 V, ratio 100) and works with a near-open secondary — it must never be short-circuited. Errors: the ratio error = (K_n − R)/R × 100% — with K_n = 20 and actual ratio 19.9, 0.50% — arises from the magnetising current, and the phase-angle error (the secondary current is not exactly 180° from the primary) from core loss and burden; both grow with burden and at low load. A Rogowski coil (air-core, output ∝ di/dt, followed by an integrator) and a Hall-effect clamp (DC to high frequency) are non-contact alternatives. Burden is the VA load on the secondary and must stay below the rated burden.
| Transformer | Secondary condition | Never |
|---|---|---|
| Current transformer | Near short circuit | Open the secondary under load |
| Potential transformer | Near open circuit | Short-circuit the secondary |
DC and single-phase. A moving-coil meter reads DC (and the average of rectified AC); extending range uses a series multiplier R_s = (V/I_m) − R_m for a voltmeter and a parallel shunt R_sh =…
True RMS, Counters, the Oscilloscope, Spectrum Analyzer and Energy Meter
True rms against average-responding. An average-responding meter rectifies, averages and scales by the sine form factor 1.11 (π/2√2): exact for a sine, wrong for anything else. A 1 V square wave (average 1 V) reads 1.11 V — +11%; a triangle (form factor 1.155) reads 4% low. A true-rms meter squares, averages and roots (thermal or computational) and is right for any shape within its crest-factor limit (crest factor = peak/rms: 1.414 sine, 1 square, 1.73 triangle, 3–10 for pulse trains). Timer/counter: frequency measurement counts input cycles in a gate time T: resolution ±1 count, so relative error = 1/(fT) — for 1 kHz and T = 1 s, 0.1%; T = 10 s, 0.01%. For low frequencies the period is timed against a clock instead: a 1 kHz period is 1 ms = 10000 counts of a 10 MHz clock, so ±1 count is 0.01%; reciprocal counting gives constant relative resolution at all frequencies. Phase = time difference between zero-crossings × 360°/period; time-interval error also includes trigger-level jitter.
Digital storage oscilloscope. The analog bandwidth (−3 dB) sets the fastest edge it can show: rise time t_r = 0.35/BW, so 100 MHz gives 3.5 ns, and a 5 ns signal edge is displayed as √(5² + 3.5²) = 6.10 ns (rule: scope bandwidth ≥ 3–5 × the signal's highest frequency). The sample rate should be ≥ 2.5–5 × the bandwidth (the Nyquist rate alone gives a crude waveform); memory depth × sample period is the capture window. A 10× probe (10 MΩ ∥ ~12 pF) reduces loading. The spectrum analyzer (swept or FFT) shows amplitude against frequency: a narrower resolution bandwidth (RBW) lowers the noise floor and separates close tones but sweeps slowly (sweep time ∝ span/RBW²). Digital voltmeter and multimeter: a 3½-digit DVM displays up to 1999 counts, so on a 200 mV range its resolution is 0.1 mV (0.05% of range), and its error is quoted as ±(% of reading + counts). Energy meter: a 1200 rev/kWh electromechanical meter turns at 1 kW × 1200 rev/h = 0.333 rev/s (20 rev/min) — a digital meter multiplies sampled voltage and current, integrates P dt, and emits pulses (e.g. 3200 imp/kWh); the energy of a load follows from N pulses ÷ pulse constant.
| Instrument | Key relation | Worked value |
|---|---|---|
| Average-responding on square wave | Reads 1.11 × average | +11% |
| Counter, gate T, frequency f | δ = 1/(fT) | 0.1% (1 kHz, 1 s) |
| Counter, period mode | δ = 1/(T_x f_clk) | 0.01% (1 kHz, 10 MHz clock) |
| Oscilloscope rise time | 0.35/BW | 3.5 ns (100 MHz) |
- Rectifies, averages, scales by 1.11
- Exact only for a sine
- 1 V square wave reads 1.11 V
- Squares, averages, roots
- Right for any shape
- Within its crest-factor limit
True rms against average-responding. An average-responding meter rectifies, averages and scales by the sine form factor 1.11 (π/2√2): exact for a sine, wrong for anything else. A 1 V square…
Key takeaways
- GUM uncertainty: Type A s/√n, Type B a/√3, combined by quadrature, expanded U = 2u_c; random errors in products add as relative errors in quadrature.
- Weights are 1/σ²; weighted means and weighted least squares trust precise points more and beat any single reading.
- Balance: Z₁Z₄ = Z₂Z₃. Wheatstone R_x = R₂R₃/R₁; Kelvin R_x = S·P/Q; Maxwell L_x = R₂R₃C₄ (low Q); Schering tan δ = ωC₄R₄; Wien f = 1/(2πRC).
- Two-wattmeter method: P = W₁ + W₂, tan φ = √3(W₁ − W₂)/(W₁ + W₂); never open a CT secondary or short a PT secondary.
- Average-responding meters misread non-sine waves; counter resolution is 1/(fT) in frequency mode and better in period mode; scope rise time is 0.35/BW.
Practice questions (24)
Attempt each one before opening the answer. Every explanation names the tempting wrong option as well as the right one, because that is where marks are lost.
P = V²/R is computed from V = 10.0 V ± 1% and R = 100 Ω ± 1% (independent random errors). The uncertainty in P in per cent (two decimals) is:
Numerical answer — type the value.
Show answer
Answer: 2.24
δP/P = √((2 × 1%)² + (1%)²) = √5 % = 2.24%; the arithmetic worst case would be 3%.Two readings of the same quantity are 10.0 ± 0.1 and 10.3 ± 0.2 (standard uncertainties). Their weighted mean, with weights 1/σ², is (three decimals):
Numerical answer — type the value.
Show answer
Answer: 10.060
Weights 100 and 25: (100 × 10.0 + 25 × 10.3)/125 = 10.060, with standard uncertainty 1/√125 = 0.0894.For a Wheatstone bridge with arms R₁ = 100 Ω, R₂ = 1 kΩ and R₃ = 235 Ω balanced against R_x (R_x = R₂R₃/R₁), R_x is:
Show answer
Answer: B — 2350 Ω
R_x = 1000 × 235/100 = 2350 Ω. Inverting the ratio would give 23.5 Ω, the usual slip.
Show the other 21 questions
A Maxwell inductance bridge is balanced with R₂ = 1 kΩ, R₃ = 500 Ω, R₄ = 2 kΩ and C₄ = 0.1 µF (L_x = R₂R₃C₄). The inductance in mH is:
Numerical answer — type the value.
Show answer
Answer: 50
L_x = 1000 × 500 × 0.1 × 10⁻⁶ = 0.05 H = 50 mH; R_x = R₂R₃/R₄ = 250 Ω.A Schering bridge is used chiefly to measure:
Show answer
Answer: B — capacitance and dielectric loss (tan δ) of insulation
With a loss-free standard capacitor in one arm, the Schering bridge gives C_x = C₂R₄/R₃ and tan δ = ωC₄R₄, the standard test for high-voltage insulation. Kelvin handles low resistance, Wien frequency, Anderson high-Q inductors.In a Q-meter test, a coil resonates with the input at 10 mV and the voltage across the tuning capacitor reads 1.2 V. The Q of the coil is:
Numerical answer — type the value.
Show answer
Answer: 120
At resonance the capacitor voltage is Q times the injected voltage: Q = 1.2/0.010 = 120.Two wattmeters measure a balanced three-phase load and read 2000 W and 1000 W. The power factor (three decimals) is:
Numerical answer — type the value.
Show answer
Answer: 0.866
tan φ = √3(2000 − 1000)/(2000 + 1000) = 0.5774, so φ = 30° and pf = cos 30° = 0.866; total power 3000 W and reactive power 1732 var.Why must the secondary of a current transformer never be left open while the primary carries current?
Show answer
Answer: D — The unopposed primary ampere-turns saturate the core and induce a dangerously high secondary voltage
In a CT the primary current is set by the line, not the transformer; with no secondary current to cancel it, the whole primary ampere-turns magnetise the core, driving it deep into saturation with a high, peaked emf.An average-responding AC meter (calibrated for sine waves, form factor 1.11) is used to measure a 1 V amplitude square wave. The percentage error in its reading is:
Numerical answer — type the value.
Show answer
Answer: 11
The average of the rectified square wave is 1 V; the meter multiplies by 1.11 and shows 1.11 V, while the true rms is 1.00 V: +11%.A frequency counter measures 1 kHz with a 1 s gate time. The ±1-count quantisation error is:
Show answer
Answer: C — 0.1%
Error = 1/(fT) = 1/(1000 × 1) = 0.1%. Timing the period against a 10 MHz clock instead gives 1 ms/100 ns = 10 000 counts, a 0.01% error, which is why low frequencies are measured in period mode.Four readings of a length are 10.2, 10.4, 10.3 and 10.1 mm. Their mean, in mm (two decimals), is ____.
Numerical answer — type the value.
Show answer
Answer: 10.25
Sum = 41.0, so the mean is 41.0/4 = 10.25 mm. Rounding to 10.3 mm too early throws away the second decimal that the average legitimately carries.Under the GUM, a Type A evaluation of standard uncertainty is one obtained by
Show answer
Answer: D — statistical analysis of a series of repeated observations
Type A comes from the scatter of repeated readings (s/√n). Data sheets, certificates, resolution and judgement are all Type B sources, so the certificate option is the tempting misfiling; the two types differ in method, not in reliability.A single-phase load draws 5 A from a 230 V supply at power factor 0.8 lagging. The real power, in W, is ____.
Numerical answer — type the value.
Show answer
Answer: 920
P = VI cos φ = 230 × 5 × 0.8 = 920 W. The product VI = 1150 VA is the apparent power, which a wattmeter does not read.The rms value of a sine wave of 10 V peak, in V (two decimals), is ____.
Numerical answer — type the value.
Show answer
Answer: 7.07
V_rms = V_p/√2 = 7.07 V. The tempting 6.37 V is the average of the rectified wave (2V_p/π), which is what an average-responding meter actually senses before its 1.11 scaling.A 200/5 A current transformer's secondary ammeter reads 3 A. The primary current, in A, is ____.
Numerical answer — type the value.
Show answer
Answer: 120
I_p = I_s × 200/5 = 3 × 40 = 120 A (ignoring ratio error). Dividing by 40 instead gives 0.075 A, the inverse ratio.A 4½-digit DVM (maximum display 19999) is on its 2 V range. Its resolution, in mV (one decimal), is ____.
Numerical answer — type the value.
Show answer
Answer: 0.1
The 2 V range displays 1.9999 V, so one count is 0.0001 V = 0.1 mV. Answering 1 mV is the 3½-digit resolution on the same range.Sixteen readings have a sample standard deviation of 0.4 units. The standard uncertainty of their mean is ____ (one decimal).
Numerical answer — type the value.
Show answer
Answer: 0.1
s/√n = 0.4/4 = 0.1. Dividing by n (0.025) overstates the benefit of averaging; the standard deviation itself (0.4) describes single readings, not the mean.A Wien bridge used for frequency measurement has R₁ = R₂ = 5 kΩ and C₁ = C₂ = 10 nF. The balance frequency, in Hz (nearest integer), is ____.
Numerical answer — type the value.
Show answer
Answer: 3183
f = 1/(2π√(R₁R₂C₁C₂)) = 1/(2π × 5 × 10⁻⁵) ≈ 3183 Hz. Leaving out 2π gives 20 000 (the angular frequency in rad/s).An energy meter has a constant of 1200 revolutions per kWh. With a steady 2 kW load for 30 minutes, its disc makes how many revolutions? ____
Numerical answer — type the value.
Show answer
Answer: 1200
Energy = 2 kW × 0.5 h = 1 kWh, so 1200 revolutions. Using 30 h instead of 0.5 h, or forgetting the time altogether (2400), are the usual slips.R is found from R = V/I with independent relative standard uncertainties of 0.5 % in V and 1.2 % in I. The combined relative standard uncertainty in R, in per cent (one decimal), is ____.
Numerical answer — type the value.
Show answer
Answer: 1.3
For a quotient, relative uncertainties combine in quadrature: √(0.5² + 1.2²) = √1.69 = 1.3 %. Adding them (1.7 %) is the worst-case limiting error, and subtracting them (0.7 %) has no basis even though division is involved.A Schering bridge balances at 50 Hz with C₄ = 0.2 µF in parallel with R₄ = 1000 Ω in the fourth arm. The dissipation factor tan δ = ωC₄R₄ of the specimen (four decimals) is ____.
Numerical answer — type the value.
Show answer
Answer: 0.0628
tan δ = 2π × 50 × 0.2 × 10⁻⁶ × 1000 = 0.0628. Using f = 50 in place of ω (0.0100) drops the 2π; the specimen's capacitance itself comes from the other balance equation, C_x = C₂R₄/R₃.A single wattmeter on a balanced three-phase load has its current coil in line R and its pressure coil across lines Y and B. It reads 1000 W. The total reactive power of the load, in VAr (nearest integer), is ____.
Numerical answer — type the value.
Show answer
Answer: 1732
V_YB is 90° from V_R, so the meter reads V_L I_L sin φ; the total reactive power is √3 V_L I_L sin φ = √3 × 1000 ≈ 1732 VAr. Answering 3000 VAr multiplies by 3 as for a per-phase wattmeter, which this is not.A pulse train switches between 0 V and 5 V with a 20 % duty cycle. Its true rms value, in V (three decimals), is ____.
Numerical answer — type the value.
Show answer
Answer: 2.236
V_rms = √(D × 5²) = 5√0.2 = 2.236 V. The average is only 1.0 V; an average-responding meter would show 1.11 V, so the 1.000 or 1.110 readings are the traps a true-rms meter avoids.Which statements about expressing measurement uncertainty are correct?
Show answer
Answer: B — Type B uncertainty may be taken from calibration certificates or data sheets.; C — A coverage factor k = 2 corresponds to roughly 95 % coverage for a normal distribution.; D — Expanded uncertainty is U = k·u_c, where u_c is the combined standard uncertainty.
U = k·u_c with k = 2 giving about 95 % for a Gaussian, and Type B covers non-statistical sources such as certificates. The standard uncertainty of a mean is s/√n, so it falls as 1/√n: quadrupling the readings only halves it, which makes the 1/n statement the trap.